23

5 Measuring

Setup of the Serial Interface

All signs have to be in capital letters and shall be sent and received as an ASCII code.

Setup:

Baud

Rate

9600

Data

bit

8

Stop

bit

1

Parity

None

Handshake XON/OFF



Pin Assignment of PLC Interface XS 5 (97123) and PLC Interface XS 9 (97103)
see Section 8.2

Selection Command

Before you start the desired dispensing cycle you have to select the appropriate
reference.

The command you have to send is “X” and the number of the reference.

For example, you send

X0

= Reference 0 is active

X1

= Reference 1 is active

X2

= Reference 2 is active

X3

= Reference 3 is active

: :

: :

X9

= Reference 9 is active

The answer you get is “ok”. This means you can start your desired dispensing cycle.

If you get an Error message, send the command again. In case of further Error
messages see Chapter 7.



Multiple choice of reference dispensing is not possible.