Measurement
Disturbances / System messages
30
31
Recall stored values:
Press S until measured value is displayed (1 second).
Scroll to next day by repressing the S key while the measured
value is displayed (5 sec). Fast scrolling by repeatedly pressing
the S key.
· Convert the displayed measured value (digits) into the BOD value with the
following table (Digits x Factor = BOD5 in mg/l):
Sample volume (ml)
Measuring range (mg/l)
Factor
432
0 - 40
1
365
0 - 80
2
250
0 - 200
5
164
0 - 400
10
97
0 - 800
20
43.5
0 - 2000
50
22.7
0 - 4000
100
Disturbances
· Measured value remains below measuring range
The display shows zero or too low a value.
The measuring equipment is not water-tight.
Check rubber quiver, screw top and bottle.
Insufficient sample pretreatment or preservation.
The temperature of the sample had not sufficiently been adjusted (< 15°C).
· Measuring range exceeded.
The measuring range chosen is too small. With very high values (> 2000 mg/l) we
recommend to predilute the sample.
Nitrification inhibitor (allylthiourea) is missing or lacking.
Errors due to procedure have not been mentioned.
System messages
Memory empty (
IF = measured value of day 1 is missing)
Change batteries (approx. every 3 years)
Value remains below measuring range (< 0 digits)
Value exceeds measuring range (> 50 digits)