Technical data, Dimensions, 2 technical data – LumaSense Technologies IMPAC IS 50-LO/GL Benutzerhandbuch

Seite 15

Advertising
background image

INFRATHERM pyrometer IS 50-LO/GL

15

2 Technical

data


Temperature range:

600 ... 1800°C

Sub-temperature range:

User-adjustable (minimum span is 51°C).

Data handling:

digital

Spectral range:

0.8 ... 1.1 µm

Fiber optic cable:

Multi-fiber light guide (Ø of 0.6 mm) with stainless steel protective coating

IR detector:

Silicon (Si)-based photo diode

Power supply:

24 V DC ± 25% (18 … 30 V DC), ripple must be less than 50 mV
power consumption max. 0.6 VA

Analog output:

4 ... 20 mA direct current, linear temperature resolution (two wire technique)

Load:

Max.: 400

Ω (18 V) to 700 Ω (24 V) to 1000 Ω (30 V)

Parameters:

Adjustable at the pyrometer:

Emissivity

ε

, switchable test output 12 mA

Also adjustable and readable via the PC interface:

Sub-temperature range (must be within the limits of the full
temperature range and must have a minimum span of 51°C).

Emissivity:

5 … 100%

set via emissivity control inside the pyrometer

Response time:

1 s (set ex works)

Accuracy (23°C):

< 1500°C: within 0.3% of measured value in °C + 1°C
> 1500°C: within 0.5% of measured value in °C

Temperature coefficient:

0.2°C per °C deviation from 23°C

Reproducibility:

0.1% of measured value in °C + 1°C

Ambient temperature:

Converter housing:

0 … 70°C

Fibre optics:

up to 250°C on optical head’s side

Storage temperature:

-20 … 70°C

Enclosure rating:

IP65 (DIN 40050)

Enclosure class:

Ι (VDE 0411)

Weight of signal processor:

600 g (approx)

Dimensions of housing:

125 mm x 80 mm x 57 mm (without plug), die-cast aluminium

Dimensions of resolving optic:

112 mm x 22 mm (l x Ø)

Operation position:

any

CE-label:

According to EU directives about electromagnetic immunity



2.1 Dimensions

Signal processor:

Optical head:

all dimensions in mm