4 electrical connection, Connection, 5 factory settings – LumaSense Technologies IN 5-DC-PL2 Ext. PL Benutzerhandbuch

Seite 7: 4electrical connection, 5factory settings

Advertising
background image

IMPAC-Pyrometer IN 5-DC-PL2

Ext.PL

7

4

Electrical connection

The series 5 pyrometers are powered by 24 V DC nominal (possible range: 10
... 30 V DC, ripple must be less than 0.5 V). When connecting the device to the
power supply ensure correct polarity. The instrument has no warm-up time,
with connection to the power the instrument operates immediately. To switch
off the instrument, interrupt the power supply or unplug the electrical connec-
tor.
Analog output and power supply are galvanically isolated.
To meet the electromagnetic requirements (EMV), a shielded connecting cable
must be used. The shield of the connecting cable is usually only connected on
the pyrometer side. If the connecting cable is extended, the extension cable
also needs to be shielded. Do not connect the shield to the power source side
(control cabinet), to avoid ground loops.


Connection:

K

A

B

C

D

J

G

F

E

H

M

L

Pin assignment

(side of male inserts)

Pin K

white

+ 24 V supply voltage

Pin A

brown

0 V supply voltage

Pin B

yellow

- linear output

Pin L

green

+ linear output

Pin H

grey

external switch for targeting light

(bridge to K)

Pin M

orange

screen

(do not connect to control cabinet)


5

Factory settings


Emissivity (Emi):

95%

Response time (t

90

):

250 ms

Analog output:

0 ... 20 mA


Advertising