22

Outputcalibrationprocedure

Whenthe

displayshows

Action/Description

м

Е

ENT

Put the instrument in setup mode and scroll through the main menu

COPn

Press

Å

or

ì

COPy

Calibrate Outputs?

Toggle

Start

Measure the output current

I=0

The output current should be 0.00 mA
(Use the keypad to adjust the value as required)

Increase Decrease Accept

I=20

The output current should be 20.00 mA

Increase Decrease Accept

Measure the output voltage

E=0

The output voltage should be 0.00 V

Increase Decrease Accept

E=10

The output voltage should be 10.00 V

Increase Decrease Accept

Save

Instrument will save values and return to normal operation (using the new values).

Note: Do not remove the power while the save message is on display.