Equipmentrequirements
• An accurate digital multimeter (accurate to 0.0 mV and ±0.1 μA)
Outputcalibrationprocedure
Whenthe
displayshows
Action/Description
м
Е
ENT
Allow fifteen minutes of powered operation (to reach a stable temperature).
Put the instrument in setup mode and scroll through the main menu
COPn
Press
*
.
COPy
Calibrate Outputs?
Toggle
Start
Measure the output current
I=0
Press
*
. The output current should be 0.00 mA
(Use the keypad to adjust the value as required)
Increase Decrease Accept
I=20
Press
*
. The output current should be 20.00 mA
Increase Decrease Accept
Measure the output voltage
E=0
Press
*
. The output voltage should be 0.00 V
Increase Decrease Accept
E=10
Press
*
. The output voltage should be 10.00 V
Increase Decrease Accept
Save
Instrument will save values and return to normal operation (using the new values).
Note: Do not remove the power while the save message is on display.