Calibration – Weidmuller UPIT - Analog to frequency converter Benutzerhandbuch

Seite 23

Advertising
background image

23

•Calibration

General
The UPI/T is factory calibrated. Once calibrated it will operate with any
input format.
Allow the instrument 1 minutes of powered operation (to reach a
stable temperature) before calibration.
During calibration allow a few seconds for signals to stabilise before
accepting the value.
Equipment requirements
• A suitable current or voltage source
• An accurate digital multimeter (accurate to 0.0 mV and ±0.1 μA)

Terminalconnectionsforinputcalibration

CalibrationStage

Signaltype

Terminal

Current (mA) inputs

Millivolt source +

2

Millivolt source −

3

Voltage (V) inputs

Voltage source +

4

Voltage source −

3