FRITSCH PULVERISETTE 7 premium line Benutzerhandbuch

Seite 45

Advertising
background image

5.

Switch on the transmitter with the On/Off button (A) and insert

the bowl as described in the operating manual of the PULVERI‐

SETTE 7 premium line in chapter 4.10 "Inserting the grinding bowl

in the grinding bowl holder"!

NOTICE!
Check the fitting of the transmission unit again after

inserting the bowl!

NOTICE!
During a grinding, during the pauses, the correct

position of the Allen screws of the transmission unit

must to be checked. If screws became loose, re‐

tighten them.

2.5 Installation of the receiver board in the P-7 premium line

DANGER!
Before starting the installation, disconnect the mains plug

and protect the device against being unintentionally

switched back on.
Secure installation work with a warning sign.

1.

The 6 screws (e) used to attach the rear panel to the housing are

undone and the rear panel is opened carefully.

EASY GTM Gas Pressure and Temperature Measuring System