1 cleaning of the sensor and changing the o-ring, 1 cleaning of the sensor and changing the o‐ring – FRITSCH PULVERISETTE 7 premium line Benutzerhandbuch
Seite 52
Advertising

d socket head screws
When inserting the locking ring (11), make sure that the two holes are
positioned over the socket head screws (d).
The lid normally remains on the bowl (10).
If the lid (9) is on the measuring/transmission unit (8), release it by hand
before cleaning. The lid (9) and the bowl (10) can be cleaned under run‐
ning water as described in the operating manual of the "PULVERISETTE 7
premium line" chapter 7.1 "Grinding elements".
2.11.1
Cleaning of the Sensor and changing the O-ring
The sensor (G) must not be subjected to any mechanical stress. It is per‐
mitted to cleaning it with a wet cloth. The O‐ring (5) can be replaced if
necessary.
EASY GTM Gas Pressure and Temperature Measuring System
‐ 52 ‐