Configuration of the interfaces
Change of the configuration of the interfaces is only possible with a VACUUBRAND vacuum control-
ler CVC 3000 (configuration of the digital interface from software version V 1.30, analog interface from
software version 1.40) or with the measurement device DCP 3000 8only digital interface from software
version V 1.00).
1. CVC 3000 or DCP 3000 is switched off.
2. Plug in only the interface to be configurated (no further components)
3. Switch on the device while key ”VENT” is pressed.
4. Select menu ”Vacuubus” by turning and pressing the selection knob.
5. The first component detected by the VACUU•BUS
TM
system is displayed (digital interface from soft-
ware version V 1.30, analog interface from software version V 1.40).
If necessary select an other component by turning the selection knob.
6. CVC 3000: Press the selection knob while key MODE is pressed.
DCP 3000: Press the selection knob while key VENT is pressed.
The inverse display of the component disappears.
7. Turn the selection knob until the designation of the desired component is displayed.
8. Press the selection knob. If the configuration was successful the configuration menu is determined
and the device restarts.
Options
For the
analog interface only the following settings are relevant: “Vacuum”, “Speed”, “Set vac” and “Set
speed”, for the
digital interface only “Valves”, “Error”, “Remote” and “End”.
Select between the alternatives by turning the selection knob to the left or the right.
Only the settings
relevant for I/O Modules are described below.
Digital interface
➨
Isol.v. 1 2 3 4: Configuration of up to 4 isolation valves
➨
Waterv. 1 2 3 4: Configuration of up to 4 coolant valves
➨
Air. V. 1 2 3 4: Configuration of up to 4 venting valves
➨
VMS 1 2 3 4: Configuration of up to 4 VMS modules
➨
Error: Configuration of a fault indicator (factory set)
➨
Remote: Configuration of a remote module (Start - Stop)
➨
End: Configuration of a valve to be opened at the end of the process
Analog interface
➨
Vacuum: Configuration output actual vacuum (factory set)
➨
Speed: Configuration output actual speed
➨
Set vac: Configuration input set vacuum and output actual vacuum
➨
Set speed: Configuration input set speed and output actual speed