A technische daten, Ichtig, Kupfertestmodul – EXFO AXS-200/610 Copper Test Module Benutzerhandbuch

Seite 189

Advertising
background image

Kupfertestmodul

179

A Technische Daten

W

ICHTIG

Änderungen an den nachstehenden technischen Daten sind ohne
Vorankündigung möglich. Die in diesem Kapitel enthaltenen
Informationen dienen nur zur Referenz. Die aktuellen technischen
Daten dieses Produkts finden Sie auf der EXFO-Website unter
www.exfo.com
.

NOTE

a. Characteristics are subject to instrument noise floor (approx

70 dBm). Levels below

70 dBm can be measured using the PSD noise test.

b. Does not include the uncertainty due to VOP.

RECEIVER CHARACTERISTICS

a

Receive frequency

200 Hz to 10 kHz, resolution 1 Hz

Receive frequency

10 kHz to 20 kHz, resolution 10 Hz

Receive frequency

20 kHz to 30 MHz, resolution 1 kHz

Frequency uncertainty (accuracy)

±0.1%

Receive level (dBm)

–90 to +10 at 100

Ω or 135 Ω,resolution 0.1 dB

–100 to +10 at 600

Ω,resolution 0.1 dB

Level uncertainty (accuracy)

±1.0 dB for 200 Hz to 20 kHz at 0 dBm
±1.0 dB for 20 kHz to 30 MHz at 0 dBm

Impedance (

Ω)

100, 135, 600 and bridging (100 k

Ω)

TRANSMITTER CHARACTERISTICS

Transmit frequency

200 Hz to 20 kHz, resolution 1 Hz steps

Transmit frequency

20 kHz to 30 MHz, resolution 1 kHz steps

Transmit level (dBm)

–10 to +10 at 100

Ω or 135 Ω

–20 to +10 at 600

Ω

Frequency accuracy

The better of ±50 ppm or ±0.5 Hz

Level uncertainty (accuracy)

±0.6 dB

200 Hz to 1 MHz

±1 dB

1 MHz to 2.2 MHz

±2 dB

2.2 MHz to 17 MHz

±3 dB

17 MHz to 30 MHz

Impedance (

Ω)

100, 135 and 600

VF NOISE MEASUREMENT

Range (dBm)

0 to –90, subject to instrument noise floor

Uncertainty (accuracy) (dB)

±1

Filters

None, 3 kHz flat, C-message, psophometric, notched and D filter (IEEE 743-1995)

Graphic results

Delay distribution and jitter histogram

VF IMPULSE NOISE

Low threshold (dBm)

0 to –40, in 1 dB steps

Mid threshold

Low threshold plus separation

High threshold

Mid threshold plus separation

Separation (dB)

1 to 6, in 1 dB steps

Dead time (ms)

125

Filters

None, 3 kHz flat, C-message, psophometric, notched and D filter (IEEE 743-1995)

Counter

Maximum 999 for each threshold

Timer

1 minute to 24 hours, default is 15 minutes

POWER INFLUENCE (NOISE TO GROUND)

Noise range (dBm)

–60 to +10

Accuracy (dB)

±1.0

Level uncertainty (accuracy) (dB)

±1.0 at –60 dBm

VF LONGITUDINAL BALANCE

Frequency (Hz)

1004

Frequency uncertainty (accuracy)(ppm)

±50

Level range (dB)

0 to 80

Level uncertainty (accuracy)(dB)

±1

TIME-DOMAIN REFLECTOMETRY (TDR)

Mode

One shot, continuous (auto-repeat) with cursor and zoom

Distance range (m)

3 to 6000 (10 ft up to 20 000 ft)

Pulse width

15 ns to 20

µ s (auto-selected in auto TDR test)

Test signals

Sine wave, compensated sine wave, half-sine wave and square wave

Amplitude

7.5 V p-p on cable, 9 V p-p open circuit

V.O.P.

0.400 to 0.999 or 120 to 299 m/

µ s

Distance uncertainty

b

(accuracy) (m)

±(0.3 + 1 % x distance) or ±(1 ft + 1 % x distance)

Units

Feet, meters and nanoseconds

Horizontal scale (m)

Automatic or 30 (100 ft), 300 (1000 ft), 600 (2000 ft), 1500 (5000 ft), 3000 (10 000 ft), 6000 (20 000 ft),
13 500 (45 000 ft) and 15 000 (50 000 ft)

LOAD COIL DETECTION

Count

Five

Plot (kHz)

Up to 10

Distance range (m)

Up to 8000 (up to 27 000 ft)

SINGLE-END FREQUENCY RESPONSE

Distance range (m)

10 to 5000 (30 ft to 16 000 ft)

Frequency range (MHz)

Up to 30

Frequency uncertainty (accuracy)(ppm)

±50

Uncertainty (accuracy) (dB)

±1.0 typical

Resolution (dB)

0.1

Horizontal scale (MHz)

ADSL2+ = 2.208, VDSL2-12 = 12, VDSL2-17 = 17.66, VDSL2-30 = 30

Vertical scale

0 to +90

Advertising