Commissioning

Field of application

The reference electrode R 503/P or R 503/D is used to
create a reference potential when measuring ion
concentrations. Together with a WTW ion sensitive
electrode of the type 500 it forms a double rod combination
electrode.

Filling in the bridge

electrolyte

1 Unscrew the white screw cap from the bridge

electrolyte bottle.

2 Pull off the seal.

3 Screw the white screw cap on the bottle again.

4 Position the spout of the white screw cap upward.

5 While holding the electrode in an upright position

push the closing ring (4) downward using your
fingers until the filling opening is uncovered (5).

6 Fill the bridge electrolyte through the filling opening

(5) using the spout.

7 Shortly press the shaft of the electrode (7) upward

against the connection head in order to wet the
ground junction (8) with bridge electrolyte.

8 When doing so refill spilled bridge electrolyte.

9 Close the bottle with bridge electrolyte solution:

Position the spout of the white screw cap downward.