What to do if ...

Measured value

unstable

Cause

Remedy

– Inner junctions not

sufficiently wetted with
bridge electrolyte

– Fill up bridge electrolyte until

the inner junctions are
covered with bridge
electrolyte

– Inner diaphragms

encrusted

– Leave the bridge electrolyte

for some hours to take effect
on the inner junctions

– Ground junction

contaminated

– Rinse ground junction with

bridge electrolyte

or

– Clean the ground junction

with methanol (see
M

AINTENANCE

)

– Cable broken

– Replace the electrode