R 503/P / R 503/D

17

ba15345e02

08/2007

Maintenance

!

Refill any used up bridge electrolyte daily.

!

In the case of increased response times carry out the
following activities:

– Clean the ground junction with methanol

– Exchange the bridge electrolyte

!

Rinse off with water any granulate material on the outside
of the electrode.

!

Clean the electrode.
To do so, open the combination electrode:

Caution

Never pull the connection cable of the electrode.
The cable might be damaged.

1 Unscrew the knurled nut (1) from the connection

head.

2 Put the connection cable into a vertical position.

3 Unscrew the connection head (2).

4 Push the connection head (2) and pressure spring (3)

over the connection cable.

5 Remove the protection cap.

6 Push the electrode through the shaft (7).