Conditioning, calibration, measurement

General information

When operating the electrode ensure that

!

the filling opening (1) for the bridge electrolyte is open

!

the inner junctions (3) are covered with bridge electrolyte

!

no air bubbles are in the bridge electrolyte

!

the depth of immersion is within the optimum range:

Before measuring

Minimum
depth of
immersion

The ground junction (4) must be
covered

Maximum
depth of
immersion

Approx. 1 cm below the fluid level (2) of
the bridge electrolyte

1

2

3

4

x 500 + R 503/P

1 Before use, condition the double rod combination

electrode according to the operating manual of the
ion sensitive electrode.

2 Remove any air bubbles in the bridge electrolyte by

slightly knocking against the shaft.

3 Calibrate according to the operating manual of the

meter and the analysis specification.