3 analog output 0/4 to 20 ma, 2 settings only via serial interface, 1 maximum / minimum value storage – LumaSense Technologies IN 5-H plus Benutzerhandbuch

Seite 18: Analog output 0/4 to 20 ma, Settings only via serial interface, Maximum / minimum value storage

Advertising
background image

IMPAC pyrometers series IN 5 plus

18

9.1.3

Analog output 0/4 to 20 mA

When the DIP switch 2 is in the ON position, the analog output will be 4 - 20 mA, in the
OFF position, the analog output will be 0 - 20 mA. The analog output has to be selected
according to the signal input of the connected instrument (controller, PLC, etc.).
If the settings directly at the pyrometer should be used the pyrometer must be switched
into offline mode (OFFL, see 8.2). To adjust the analog output via software see section
10, Settings via interface and software).


9.2

Settings only via serial interface

9.2.1

Maximum / minimum value storage

If the maximum value storage is switched on always the highest last temperature value
will be displayed and stored. The minimum value storage saves the lowest measure-
ment taken during a reading (e.g. practical when monitoring cooled goods such as fro-
zen foods). You The storage has to be cleared at regular intervals for exchanging by a
new and actual value.
This feature is particularly useful when fluctuating object temperatures cause the display
or the analog outputs to change too rapidly, or the pyrometer is not constantly viewing
an object to be measured. In addition, it may also be beneficial to periodically delete and
reset the stored maximum values.

The following settings are possible:
off:

The max. value storage is switched off and only momentary values are measured.

0.01...25 s: If any clear time between 0.01 s and 25 s is set, the maximum value is estimated and held in

double storage mode. After the entered time the storage will be deleted.

extern:

The external clearing can be activated and used within an own software (see section 14, Data
format UPP

®

) or via an external contact (for connection see 4.1 Pin assignment for the con-

nector on the back side of the pyrometer). In this case, the storage operates only in single
storage
, because only a single deletion mechanism is used.

auto:

The “auto” mode is used for discontinuous measuring tasks. For example objects are trans-
ported on a conveyer belt and pass the measuring beam of the pyrometer only for a few sec-
onds. Here the maximum value for each object has to be indicated. In this mode the maximum
value is stored until a new hot object appears in the measuring beam. The temperature which
has to be recognized as “hot“ is defined by the low limit of the adjusted sub range. The stored
maximum value will be deleted when the temperature of the new hot object exceeds the low
limit “from“ of the sub range by 1% or at least 2°C. If a lower limit is not entered, the maximum
value storage will be deleted whenever the lower level of the full measuring range has been ex-
ceeded.
The minimum value function operates in the opposite manner. The minimum value stored will
be deleted when the upper limit of the set range is below it by 1% (at least 2°C). If an upper limit
is not entered, the minimum value stored will be deleted whenever it drops below the upper
level of the full measuring range.

Operation note: dependent on the settings the maximum value storage either works in single storage mode
or in double storage mode:

Single storage: the single storage is used when you want to reset the stored value using an external im-
pulse via one contact closure from an external relay (i.e. between two measured objects). The relay contact
is connected directly to the pyrometer between pins J and K. This mode allows a new value to be estab-
lished, after each impulse from the reset signal.

double storage: when entering the reset intervals via push buttons or PC interface the double storage is
automatically selected. This mode utilizes two memories in which the highest measured value is held and is
deleted alternately in the time interval set (clear time). The other memory retains the maximum value
throughout the next time interval. The disadvantages of fluctuations in the display with the clock frequency
are thereby eliminated.

Note:

The maximum value storage follows the function of adjustment of exposure time.
This results in:

• clear time ≤ the adjusted response time is useless

• clear times must be at least 3 times longer than the response time

• only maxima with full maximum value can be recorded, which

appear at least 3 times longer than the response time.

Settings:

4...20 mA 0...20 mA

Offli

Settings via

interface:

OFF

0.1 s

.

.

.

25 s

extern

auto

Advertising
Dieses Handbuch ist für die folgenden Produkte bezogen werden: