IMPAC pyrometers series IN 5 plus

6

IN 5-L plus
T=0...300°C:

0.6% of reading in °C or 2°C (T

amb

= 15...30°C)

1)

1% of reading in °C or 3°C (T

amb

= 0...15 or 30...63°C)

2)

T=300...900°C:

1% of reading in °C (T

amb

= 15...30°C)

1.5 % of reading in °C (T

amb

= 0…15 or 30…63°C)

IN 5/5 plus
T=100...1300°C : 0.6% of reading in °C or 2°C (T

amb

= 15...30°C)

1)

1% of reading in °C or 1.5°C (T

amb

= 0...15 or 30...63°C)

1)

T=1300...1800°C: 0.8% of reading in °C (T

amb

= 15...30°C)

1)

1.2% of reading in °C (T

amb

= 0...15 or 30...63°C)

1)

T=1800...2500°C: 1% of reading in °C (T

amb

= 15...30°C)

1.4 % of reading in °C (T

amb

= 0...15 or 30...63°C)

IN 5/5-H plus
T=200...1300°C : 0.6% of reading in °C or 2°C (T

amb

= 15...30°C)

1)

1% of reading in °C or 1.5°C (T

amb

= 0...15 or 30...63°C)

1)

T=1300...1800°C: 0.8% of reading in °C (T

amb

= 15...30°C)

1)

1.2% of reading in °C (T

amb

= 0...15 or 30...63°C)

1)

T=1800...2500°C: 1% of reading in °C (T

amb

= 15...30°C)

1.4 % of reading in °C (T

amb

= 0...15 or 30...63°C)

IN 5/5-L plus
T=200...1300°C : 0.8% of reading in °C or 3°C (T

amb

= 15...30°C)

1)

1% of reading in °C or 4°C (T

amb

= 0...15 or 30...63°C)

2)

T=1300...1800°C: 0.8% of reading in °C (T

amb

= 15...30°C)

1)

1.2% of reading in °C (T

amb

= 0...15 or 30...63°C)

1)

T=1800...2500°C: 1% of reading in °C (T

amb

= 15...30°C)

1.4 % of reading in °C (T

amb

= 0...15 or 30...63°C)

IN 5/4 plus
T=300...1300°C : 0.6% of reading in °C or 4°C (T

amb

= 15...30°C)

1)

1% of reading in °C or 6°C (T

amb

= 0...15 or 30...63°C)

1)

T=1300...1800°C: 0.8% of reading in °C (T

amb

= 15...30°C)

∗)

1.2% of reading in °C (T

amb

= 0...15 or 30...63°C)

1)

T=1800...2500°C: 1% of reading in °C (T

amb

= 15...30°C)

1.4 % of reading in °C (T

amb

= 0...15 or 30...63°C)

Repeatability:

(

ε

= 1, t

90

= 1 s)

0.3 % of measured value in °C or 0.6 °C

(whichever value is greater. The instrument must be at a constant ambient tempera-
ture for a minimum of 15 or 30 minutes;

1)

and

2)

as stated above)

Noise Equivalent Temperature
Difference (NETD):
(

ε = 1, σ = 1, T

U

= 23°C)

IN 5 plus: at

t

90

= 80 ms:

0.2°C (at 23°C meas. temperature)

at t

90

= 1 s:

0.05°C (at 23°C meas. temperature)

IN 5-H plus: at

t

90

= 10 ms:

0.7°C (at 23°C meas. temperature)

at t

90

= 1 s:

0.1°C (at 23°C meas. temperature)

IN 5-L plus:

at t

90

= 180 ms:

0.3 °C (at 23°C meas. temperature)

at t

90

= 180 ms:

0.2 °C (at 200°C meas. temperature)

at t

90

= 1 s:

0.15°C (at 23°C meas. temperature)

at t

90

= 1 s:

0.1°C (at 200°C meas. temperature)

IN 5/5 plus:

at t

90

= 80 ms:

0.7°C (at 110°C meas. temperature)

at t

90

= 1 s:

0.4°C (at 110°C meas. temperature)

IN 5/5-H plus: at t

90

= 10 ms:

0.5°C (at 500°C meas. temperature)

at t

90

= 10 ms:

0.3°C (at 1100°C meas. temperature)

IN 5/5-L plus: at t

90

= 80 ms:

1.5°C (at 300°C meas. temperature)

at t

90

= 80 ms:

0.6°C (at 500°C meas. temperature)

at t

90

= 1 s:

0.4°C (at 300°C meas. temperature)

at t

90

= 1 s:

0.2°C (at 500°C meas. temperature)

IN 5/4 plus:

at t

90

= 80 ms:

0.6°C (at 500°C meas. temperature)

at t

90

= 80 ms:

0.2°C (at 1100°C meas. temperature)

Exposure time t

90

:

IN 5 plus, IN 5/5 plus, IN 5/5-L plus, IN 5/4 plus:
0.08 s; adjustable in the pyrometer: 0.5 s; 1 s; 2 s; 5 s,

adjustable via interface: 0.5 s; 1 s; 2 s; 5 s; 10 s; 30 s

1

Whichever value is greater. The instrument must be at a constant ambient temperature for a minimum of 15 minutes
and has to be connected to the power supply.

2

Whichever value is greater. The instrument must be at a constant ambient temperature for a minimum of 30 minutes
and has to be connected to the power supply.