14 data format upp® (universal pyrometer protocol), Data format upp, Universal pyrometer protocol) – LumaSense Technologies IN 5-H plus Benutzerhandbuch

Seite 26: Ction 14

Advertising
background image

IMPAC pyrometers series IN 5 plus

26

14

Data format UPP

®

(Universal Pyrometer Protocol)


Via interface and a suitable communication software or via “Test” function of the InfraWin software (see 10.7
Basic settings

Æ Test) commands can be exchanged directly with the pyrometer.

Note:

The ONL/OFFL switch must be in the ONL position before adjusting the
following parameters via the software (see 8.2).


The data exchange occurs in ASCII format with the following transmission parameters:
The data format is: 8 data bits, 1 stop bit, even parity (8,1,e)
The device responds to the entry of a command with: output (e.g. the measuring value) + CR (Carriage Re-
turn, ASCII 13), to pure entry commands with “ok” + CR.
Every command starts with the 2-digit device address AA (e.g. “00”).
This is followed by 2 small command letters (e.g. “em” for emissivity), followed, if necessary for that com-
mand, by the ASCII parameter “X” and finished with CR. If this parameter “X” is omitted, then the device
answers with the current set parameter.
A „?“ after the small command letters answers with the respective settings (only at setting commands, not at
enquiry commands).

Example:

Entry: “00em“ +

<CR>

The emissivity setting (

ε

) of the device with the address 00 is returned

Answer: “0970“ +

<CR> means Emissivity = 97.0%


Description

Command

Parameters

Reading measuring value:

AAms

Output: YYYYY

5-digit decimal, in

1

/

10

°C or. °F

88880 = temp. overflow

02563 = 256.3 °C or °F

-0170 = -17.0 °C or °F

Emissivity:

AAemXX
AAemXXXX

XX = 20 to 99%, 00 = 100% (decimal)
XXXX = 0200 to 1000 in ‰

Exposure time t

90

:

AAezX

X = 0 to 6 (decimal)

0 = intrinsic time constant of the device

1 = 0.5 s

3 = 2 s

5 = 10 s

2 = 1 s

4 = 5 s

6 = 30 s

Analog output

1)

:

AAasX

X = 0;1 0 = 0 - 20 mA

1 = 4 - 20 mA

Clear time of maximum /
minimum value storage:

AAlzX

X = 0 to 8 (decimal)

0 = maximum/minimum value storage off

1 = 0.10 s

5 = 5.00 s

2 = 0.25 s

6 = 25.00 s

3 = 0.50 s

7 = externally deleted

4 = 1.00 s

8 = automatically cleared

External clearing:

AAlx

Simulation of an external clearing contact

Changing °C / °F

1)

:

AAfhX

X = 0;1 (decimal)

0 = Output Celsius

1 = Output Fahrenheit

Reading basic temperature
range:

AAmb

Output: YYYYZZZZ

8-digit hex in °C or °F

YYYY = beginning of temperature range

ZZZZ = end of temperature range

Reading sub range:

AAme

as with mb

Repeatedly reading
measuring values:

AAmsXXX

XXX = 000 to 999
XXX = No. of measuring values

Laser targeting light:

AAlaX

X = 0;1 (decimal)

0 = switch off pilot light

1 = switch on pilot light