IMPAC pyrometers series IN 5 plus

27

Error status:

AAfs

Output 1 Byte hex

Bit 0 = 1: EEPROM error

Bit 1 = 1: Watch dog Reset

Bit 2 = 1: Under voltage reset

Reading parameter:

AApa

Output 11-digit decimal:

Digit 1 and 2 (20 to 99 or 00): emissivity

Digit 3 (0 to 6): t

90

(exposure time)

Digit 4 (0 to 8): t

CL

(storage delete mode)

Digit 5 (0 to 1): analog output

Digits 6 and 7 (00 to 98): temperature

Digits 8 and 9 (00 to 31): address

Digit 10 (0 to 4): baud rate

Digit 11 always 0

Device address

1)

:

AAgaXX

XX: Device address decimal

00 to 31 variable device addresses

Baud rate:

AAbrX

X = 0 to 4

0: 1200 Bd

2: 4800 Bd

4: 19200 Bd

1: 2400 Bd

3: 9600 Bd

Internal pyrometer
temperature:

AAgt

Output: XX (dec. 00 to 98, in °C)

Max. internal pyrometer
temperature:

AAtm

Output: XX (dec. 00 to 98, in °C)

Reset

1)

:

AAre

Reset device

Command delay:

AAtwXX

XX = 00 to 20 relative delay value

Reading ambient
temperature:

AAut

Output: stored value, 4-digit hex
e.g. 0258 corresponds to 600 degrees

Entering ambient
temperature:

AAutXXXX

XXXX = value of ambient temperature, 4-digit, hex
XXXX e.g. FFEC corresponds to –20 degrees
- 99

dez

= FF9D

hex

means: automatic, no manual

compensation

Marginal values for ambient
temperature:

AAut?

Output: marginal values for the entry, 2 x 4-digit, hex
e.g. FF9D0384 corresponds to -99 to 900 degrees

Reading maximum/minimum
values:

AAmi

Output: 0 or 1
0 = maximum value, 1 = minimum value

Setting maximum/minimum
values:

AAmiX

X = desired setting
0 = maximum value, 1 = minimum value

Reading marginal values:

AAmi?

Output: marginal value for entry, always 01

Serial number:

AAsn

Output: XXXXX (5-digit decimal)

Device model / software
version:

AAve

Output: XXYYZZ (6-digit decimal)

XX = 70 (IN 5 plus)

71 (IN 5/5 plus)

YY = month of the software version

ZZ = year of the software version

Note: the letter “l” means the lower case letter of “L”.

1)

After entering these commands the device carries out an automatic reset. The device needs
approx. 150 ms before it is ready to use and work with the changed settings.


Additional instruction for the RS485 interface:
Requirements to the master system during half-duplex operation:

1. After an inquiry, the bus should be switched into a transmission time of 3 bits

(some older interfaces are not fast enough for this).

2. The pyrometer's response will follow after 5 ms at latest.
3. If there is no response, there is a parity or syntax error and the inquiry has to be repeated.