WIKA TC12-M Benutzerhandbuch

Seite 27

Advertising
background image

GB

7. Information on mounting and operation in hazardous areas

14083458.01 09/2013 GB/D

WIKA operating instructions models TR12, TC12 (Ex i)

27

The values in brackets apply to special designs. These sensors are manufactured using

special sealing compounds. Moreover, they are equipped with connection heads made of

stainless steel and cable glands for the low-temperature range.

According to approval, these thermometers are suitable for the temperature classes

T6 ... T3. This applys for instruments without built-in transmitter and/or digital displays.

Thermometers with transmitters can be used for temperature classes T6 ... T4 and are

marked accordingly. The use of equipment for applications with lower temperature classes

(for example T2) than marked is admissible. Make sure the maximal ambient temperature

for the safe use of the instrument is not exceeded.

7.3 Overview of the temperature zones

FF-00147.00

Tx12

Tx12

Option: with built-in

transmitter e.g. T32

Tx12-A

Tx12-A

T3

T3

Pr

ocess connection

Pr

ocess connection

1)

Permissible temperatures at

T1: (-50) -40 °C < T

a

< +300 °C

T3: (-50) -40 °C < T

a

< +150 °C

T4: (-50) -40 °C < T

a

< +100 °C

T4

Transmitter

: (-50) -40 °C < T

a

< +85 °C

1) Temperature zone undefined

T4

T4

Transmitter

T1

T1

1)

Advertising
Dieses Handbuch ist für die folgenden Produkte bezogen werden: