9. Calculation examples ... / 10. Maintenance and cleaning

14083458.01 09/2013 GB/D

WIKA operating instructions models TR12, TC12 (Ex i)

32

Additional information:

Temperature class for T3 = 200 °C

Safety factor for type-tested instruments (for T6 to T3) = 5 K

Safety factor for type-tested instruments (for T1 to T2) = 10 K

Safety factor for applications for instrument category 1 (zone 0) = 80 % finds no application

here.

9.2 Example calculation for variant 2 with TC sensor

Under the same conditions it gives a lower value for the self-warming, since the supplied

power is not only converted at the sensor tip, but rather over the entire length of the

measuring insert.

Thermal resistance [Rth in K/W] from table = 3 K/W

Self-heating: 0.8 W x 3 K/W = 2.4 K

T

max

= T

M

+ self-heating: 150 °C + 2.4 °C = 152.4 °C

As safety margin for type-examined instruments (for T6 to T3), an additional 5 °C must be

subtracted from the 200 °C; hence 195 °C would be permissible. This means that in this

case temperature class T3 is not exceeded.

This example shows that the self-heating is almost negligible here.

10. Maintenance and cleaning

10.1 Maintenance

The thermometers described here are maintenance-free.
Repairs must only be carried out by the manufacturer.

10.2 Cleaning

CAUTION!

Prior to cleaning, disconnect the electrical connections properly.

Clean the instrument with a moist cloth.

Electrical connections must not come into contact with moisture.

Wash or clean the dismounted instrument before returning it, in order to

protect persons and the environment from exposure to residual media.

Residual media in the dismounted instrument can result in a risk to persons,

the environment and equipment. Take sufficient precautionary measures.

For information on returning the instrument see chapter 12.2 “Return”.