Ambient conditions

Ambient temperature range

M12 x 1, 4-pin circular connector

-50 ... +85 °C (-58 ... +185 °F)

Directly connected cable

-20 ... +80 °C (-4 ... +176 °F)

Storage temperature range

M12 x 1, 4-pin circular connector

-40 ... +85 °C (-40 ... +185 °F)

Directly connected cable

-20 ... +80 °C (-4 ... +176 °F)

Climate class per IEC 60654-1

M12 x 1, 4-pin circular connector

Cx (-50 ... +85 °C or -58 ... +185 °F,

5 ... 95 % relative air humidity)

Directly connected cable

Cx (-20 ... +80 °C or -4 ... +176 °F,

5 ... 95 % relative air humidity)

Maximum permissible humidity

per IEC 60068-2-30 var. 2

relative humidity 100 %,

condensation allowed

Maximum operating pressure

10) 11)

140 bar with 3 mm sensor diameter

270 bar with 6 mm sensor diameter

Vibration per IEC 60751

10 ... 2,000 Hz, 20 g

10)

Shock

IEC 60068-2-27

Salt fog

IEC 60068-2-11

10) Dependent on the instrument version
11)

Reduced operating pressure when using a compression fitting:

Stainless steel

PTFE

max. 100 bar

max. 8 bar