Main Window - Description of the Menu Items

28

Fig. 11: Sieve query AS 200 jet
This question is used as the basis for deciding on the following procedure. If the

answer is “yes”, the sieve analysis proceeds in accordance with what is known as

the “Swiss procedure“, whereby a complete representative sub-sample of the

material to be sieved is used for each mesh size. Ideally all sub-samples should

have the same weight. If the answer is “no”, the sieving analysis proceeds in

accordance with the “standard process”, whereby the oversized grains in the

respective sieving must be quantitatively transferred to the following sieve and

sieved further. Accordingly the quantity of the material to be sieved is only entered

once and after that the software enters it automatically for the following sieving

process after the deduction of one sieving run.
Information appears at the top of the window about the settings made in the sieving

task for the balance and sieving machine. When automatic weight transfer is set,

the weight is transferred automatically by actuating the weighing process on the

balance.
In the following the tare weights of the sieves, the cumulated weight of the sample

and the individual sieve weights after the sieving must be transferred or entered

into the corresponding fields. The value at the balance is always transferred into

the input box where the cursor is.
All fields are greyed out in the measuring mode because no manual changes may

be permitted here. The active field here is marked with three asterisks. In this

window the weight (only) is entered by data transfer from the balance.

Input of the tare weights:

If a sieving was conducted already for the selected sieve stack file, the fields are

filled with the last tare weights. The user has the option of accepting these now. In

that case the cursor must simply be positioned at the ““total weight” input field. If

you wish to update the tare weights, these are entered or transferred, beginning

with “bottom”, whereby the cursor advances automatically to the next field.

Input of the sieve load:

The total weight of the sample must be transferred / entered into this field. Once

the sieve load has been inputted, it is no longer possible to change the tare

weights. The prompt for starting sieving comes then.

Conducting the sieving:

Once the sieve load has been inputted, the program prompts the execution of the

sieving process. It can also be started by pressing the Sieve key. A precondition for

Advertising