29

this is that the tare weights and the sieve load are inputted completely depending

on the selected sieving machine.

Input of the sieve weights after finishing sieving with AS 200 control, AS 200
Tap, AS 300 control, AS 400

control or AS 450 control:

At the end of the sieving process the cursor moves to the “sieve weight” input field

for the coarsest sieve. Now one after another, from coarse to fine, the sieving back

weights must be transferred or entered with the sample material remaining on the

sieves. If the value transferred is lighter than the corresponding tare weight, a fault

alarm is given.

Input of the sieve weights after the end of sieving with AS 200 jet:

At the end of the respective sieving process the cursor moves to the “sieve weight”

input field for the respective sieve, from fine to coarse.
After the input of each value the calculated fractions are displayed.
Once all values have been inputted, the evaluation can be started by clicking on

the “Evaluation “ button, whereby the files defined in the sieving task (*.rdf; optional

*.xle, *.xld, *.ccg) are written and opened evaluation windows are updated. When

sieving is conducted with the AS 200 jet sieving machine in accordance with the

“Swiss procedure”, information appears once the “evaluation” key is pressed that

the calculation is based on an averaged weight of all fractions used for sieving.
In addition, a check is run through for agreement between the total sieving residue

and the inputted gross weight. A warning is given if the differences are > 2%.

Pos: 5.30 /00005 Ü berschriften/1.1.1 Ü berschriften/111 N eue Siebaufgabe @ 2\mod_1260180940116_9.doc @ 17750 @ 3 @ 1

5.3.2

New sieve task

Pos: 5.31 /00020 BDA Softwar e/20010 EasySieve Kapitelsamml ung/0004 EasySi eve H auptfensters/04151 EasySieve M odul N eue Siebaufgabe @ 2\mod_1260181189014_9.doc @ 17771 @ @ 1

Use this command for the quick generation of a new sieving task. All setting

windows required to configure the sieving task are run through. No pre-

initialisations by other sieving tasks are used.
By inputting a “file name”, you can quickly pre-initialise the name of the sieving

task, the sieve stack file, the file directory for the sieving task and the names of the

individual sieving files.
Once all sieving parameters have been set, you can start the sieving if you wish.

Pos: 5.32 /00005 Ü berschriften/1.1.1 Ü berschriften/111 Siebaufg abe bearbeiten @ 2\mod_1260180997079_9.doc @ 17757 @ 3 @ 1

5.3.3

Edit Sieve Task

Pos: 5.33 /00020 BDA Softwar e/20010 EasySieve Kapitelsamml ung/0004 EasySi eve H auptfensters/04152 EasySieve M odul Siebaufg abe bear beiten @ 2\mod_1260181288491_9.doc @ 17778 @ @ 1

The currently loaded and most recently executed sieving task is opened for editing.

All setting windows required for editing the sieving task are run through (“Next”

key).

Pos: 5.34 /00005 Ü berschriften/1.1.1 Ü berschriften/111 Siebaufg abe l aden @ 2\mod_1260181075697_9.doc @ 17764 @ 3 @ 1

5.3.4

Loading the Sieving Task

Pos: 5.35 /00020 BDA Softwar e/20010 EasySieve Kapitelsamml ung/0004 EasySi eve H auptfensters/04153 EasySieve M odul Siebaufg abe l aden @ 2\mod_1260181755844_9.doc @ 17785 @ @ 1

From a list of all previously set up sieving tasks a set of parameters can be loaded

for the sieving to be conducted. After confirmation by pressing the “start …” key,

the “start sieve analysis” window appears in which the header data of relevance to

the following sieving can be edited. When the “Edit …” key is actuated, the “Sieving

Parameters” window opens and the pending sieve analysis can be edited in this

window.

Fig. 12: Load sieve task

Pos: 5.36 /00005 Ü berschriften/1.1.1 Ü berschriften/111 Siebaufg abe speichern @ 2\mod_1254143031445_9.doc @ 16416 @ 3 @ 1