Metrohm 785 DMP Titrino Benutzerhandbuch

Seite 107

Advertising
background image

3.2 Remote control commands

785 DMP Titrino

103

.MinRate: Minimum titration rate in ul/min.

3.2.2.30. Mode.Parameter.SET1.Stop.Type drift, time

Mode.Parameter.SET1.Stop.Drift 1...20...999

Mode.Parameter.SET1.Stop.Time 0...10...999, inf

Mode.Parameter.SET1.Stop.StopT 0...99

999, OFF

Parameters for SET: Type and size of the stop criterion of the titration.

.Type: Type of stop criterion after stop drift or switch-off delay time.

.Drift: Stop drift in ul/min. Applies when "drift" has been selected.

.Time: Switch-off delay time in s. Applies when "time" has been selected. "inf"

means infinite.

.StopT: Stop time in s. Applies when "time" has been selected and the value of

.Time is set to "inf.".

3.2.2.31. Mode.Parameter.TitrPara.Direction +,

-, auto

Parameters for SET: Titration direction.

"auto" means the titration direction is determined automatically by the instru-

ment. If 2 EP's have been set in a SET titration, the titration direction is given

by the two EP's. The entry of the titration direction is then invalid.

3.2.2.32. Mode.Parameter.TitrPara.XPause

0...999 999

Parameter for SET: Pause time in s. Runs before dosing the start volume.

3.2.2.33. Mode.Parameter.TitrPara.ExtrT

0...999 999

Parameter for SET: Extraction time in s.

3.2.2.34. Mode.Parameter.TitrPara.TDelta

1...2...999 999

Parameter for SET: Time interval in s for the entry of a measurement point in

the list of measured points.

3.2.2.35. Mode.Parameter.Presel.Cond ON, OFF

Mode.Parameter.Presel.DriftDisp ON, OFF

Mode.Parameter.Presel.DCor.Type auto,

man., OFF

Mode.Parameter.Presel.DCor.Value 0.0...99.9

Parameters for SET:

.Cond: Conditioning ON/OFF

.DriftDisp: Drift display during conditioning ON/OFF.

.DCor.Type: Type of drift take-over for the drift correction. auto: Take-over of

the drift value at start.

.DCor.Value: Drift value for the manual drift correction.

3.2.2.36. Mode.Parameter.Measuring.SignalDrift

pH, U, Ipol, T:

0.5...999, OFF

Upol: 0.05...99.9, OFF

Mode.Parameter.Measuring.UnitSigDrift read

only

Mode.Parameter.Measuring.EquTime 0...9999, OFF