3.2 Remote control commands

785 DMP Titrino

123

Setup.AutoInfo.C.R2 ON, OFF

Setup.AutoInfo.I ON, OFF

Setup.AutoInfo.O ON, OFF

ON means that the Titrino reports automatically the moment the correspond-

ing change occurs.

.Status:

Global switch for all set AutoInfo.

.P

PowerOn: Simulation of power on (3.2.2.127). Not from mains.

Messages from node .T, Titrator:

.T.R

Ready: Status 'Ready' has been reached.

.T.G

Go: Instrument has been started.

.T.GC

GoCommand: Instrument (or submethod in TIP) has received a go

command.

.T.S

Stop: Status 'Stop' has been reached.

.T.B

Begin of sequence (or submethod).

.T.F

Final: End of determination (or submethod), the final steps will be

carried out.

.T.E

Error. Message together with error number, see page 73ff.

.T.H

Hold: Status 'Hold' has been reached.

.T.C

Continue: Continue after hold.

.T.O

Conditioning OK: EP reached (in SET with conditioning).

.T.N

Conditioning Not OK: EP not reached (in SET with conditioning).

.T.Re

Request: In the inquiry of an identification or the sample size after

start of titration.

.T.Si

SiloEmpty: Silo empty, i.e. the last line has been removed from

the silo memory.

.T.M

MeasList: Entry in the measuring point list (with DET, MET, SET,

MEAS).

.T.EP

EPList: Entry into EP list (with DET, MET, SET)

.T.RC

Results have been recalculated.

Messages from node .C, Comport:

.C.B1

COM1: A report is outputted on COM1. During this time, COM2

will be blocked. COM2 is generally blocked, if COM1 is busy.

.C.R1

COM1 is ready again. (Comes also when you <QUIT> an error.)

.C.B2, .R2 Identical for COM2.

Messages for changings in the I/O lines. If the changings are made simulta-

neously, there is 1 message. Pulses receive 2 messages: one message each

for line active and inactive.

.I

Input: Change of an input line.

.O

Output: Change of an output line (except 7, pin 2, for recorder

pulses).

If a change occurs that requires a message, the Titrino sends space (ASCII

32) and ! as an introducer. This is followed by the name of the device (see

3.2.2.79). Special ASCII characters in the device name are ignored. If no de-

vice name has been entered, only ! is sent. Finally the Titrino sends the infor-

mation which node has triggered the message.

Example: !John".T.Si": The message was triggered from instrument "John",

node .T.Si

3.2.2.126. Setup.Graphics.Grid ON, OFF

Setup.Graphics.Frame ON, OFF

Setup.Graphics.Scale Full, Auto

Setup.Graphics.Recorder.Right 0.2...0.5...1.00

Advertising