3.2. Remote control commands

785 DMP Titrino

124

Setup.Graphics.Recorder.Feed 0.01...0.05...1.00

Change in the appearance and the format of the curve for the output. The set-

tings are valid for both Titrino COM ports.

.Grid:

On/off switching of grid over curve.

.Frame:

On/off switching of frame surrounding the curve. If grid and frame

are switched off, the curve is printed faster as the print head does

not have to move to the end of the paper.

.Scale:

Type of scaling of the measured value axis: Full means that the

scale runs from the smallest up to the greatest measured point.

With auto, the smallest measured value is taken and the next

smaller tick defines the beginning of the scale; the next greater

tick to the greatest measured value is the end of the scale.

.Right:

Relative specification of the width of the output medium (e.g. pa-

per width) for the length of the measured value axis. 1 means the

measured value axis is plotted over the entire width of the paper

(largest possible width). In extreme cases, the writing of the right

tick may lie outside.

.Feed:

Length of the volume axis referred to the burette cylinder volume,

V(B) per cm (0.1 means, e.g. 1 mL/cm with a 10 mL Exchange

Unit). Depending on the printer, the measure in cm may not al-

ways be correct.

3.2.2.127. Setup.PowerOn $G

Simulation of 'power on'. The device has the same status as after power on:

The cylinder is filled, error messages deleted and the current sample number

set to 0. The method last used is ready for operation.

3.2.2.128. Setup.Initialise $G

Setup.Initialise.Select ActMeth, Silo, Calib, Config,

Assembly, Setup, All

Setting of default values for the following areas:

ActMeth: Current method. Parameters, calculations, and assignments for

the data output, operands C01...C19.

Silo:

The silo memory is deleted. Same function as delete entire silo.

Calib:

pH calibration data for all measuring inputs.

Config:

All values under &Config.

Assembly: All values under &Assembly.

Setup:

All values under &Setup.

All:

Values of the entire tree (except silo and method memory).

The action must be triggered with &Setup.Initalise $G.

3.2.2.129. Setup.RamInit $G

Initializes instrument, see page 148. All parameters are set to their default

value and error messages are cleared. The user and silo memories will be de-

leted. The user memory contains the default user methods from Metrohm.

3.2.2.130. Setup.InstrNo $G

Setup.InstrNo.Value serial

number, 8 ASCII characters

Instrument identification for report output.