Installation
9
ba75694e02
10/2009
1.2
Recommended fields of application
Measurements in ultrapure water in flow-through or immersion operation,
depending on the mounted outside electrode.
2
Installation
2.1
Changing between the flow-through cell and immersion
cell
The outside electrode is connected to the shaft via a screw connection and
can be removed and replaced as necessary without using any tools. When
mounting the electrode tighten the screw connection by hand as far as it will
go.
2.2
Use as flow-through cell
To avoid the collection of air bubbles in the electrode area, the flow-through
cell should be positioned so that the outlet opening is at the highest point. An
incline of the sensor by up to 60° supports the removal of air bubbles (see
following figure).
3
Cleaning
Caution
Prior to cleaning, disconnect the sensor from the meter.
Outlet
Maximum
incline 60°
Outlet points
upward
Inlet