Xylem LF213T_LF213T-ID Benutzerhandbuch
Seite 19

LF213T
Technical data
13
ba75694e02
10/2009
Measurement
conditions
Conductivity measuring
range
0.0001
μ
S/cm ... 30
μ
S/cm
Temperature range
-5 ... 80 °C (100 °C) / 23 ... 176 °F (212 °F)
Max. allowed overpressure
2 x 10
5
Pa (2 bar)
Minimum immersion depth
in immersion cell operation
40 mm
Maximum immersion depth
(at temperature)
Total sensor+cable (up to 80 °C / 176 °F)
Sensor shaft only = 120 mm (up to 100 °C /
212 °F)
Operating position
Immersion cell:
Any
Flow-through cell: Vertical or inclined by up
to approx. 60°, inlet
pointing downward
Storage conditions
Recommended storing
method
in air
Storage temperature
0 ... 50 °C / 32 ... 122 °F
Characteristic data
on delivery
Temperature responding
behavior
t
99
(99 % of the final value display after)
< 100 s
Precision of the temperature
sensor
± 0.2 K