15 reference numbers, 1 reference numbers of instruments, Reference numbers – LumaSense Technologies ISR 12-LO Benutzerhandbuch
Seite 27: Reference numbers of instruments, Tion 15

IMPAC pyrometers ISR 12-LO · IGAR 12-LO
26
Reading parameters:
AApa
Output decimal 11-digit:
Digit 1 und 2 (10...99 or 00): (emissivity)
Digit 3 (0 ... 6): t
90
(exposure time)
Digit 4 (0 ... 8): t
CL
(clear time of max. storage)
Digit 5 (0 ... 1): analog output
Digit 6 und 7: (00 ... 98): pyrometer’s temperature
Digit 8 und 9 (00 ... 97): address
Digit 10 (0 ... 6 or 8): baud rate
Digit 11 (0): (0 or 1): keyboard active or locked
Digit 12 ... 15 (0800 ... 1200): emissivity slope
Autoreset:
AAre
Activating auto-reset, initializes the pyrometer new and
resets the instrument to default settings
Reference number:
AAbn
Output: XXXXXX (hex 6-digit)
Serial number:
AAsn
Output: XXXXX (5-digit, decimal)
Device type /
software version:
AAve
Output: XXYYZZ (6-digit, decimal)
XX = 63 (ISR 12-LO or IGAR 12-LO)
YY = month of software version
ZZ = year of software version
Software version
in detail:
AAvs
tt.mm.yy XX.YY
tt = day; mm = month; yy = year; XX.YY = software version
Device type:
AAna
Output: “ISR 12-LO“ or “IGAR 12-LO (16 ASCII-characters)
Note: the letter “l” means the lower case letter of “L”.
15
Reference numbers
15.1
Reference numbers of instruments
3 855 100
ISR 12-LO
MB 13:
600 ... 1300°C
3 855 110
ISR 12-LO
MB 18:
750 ... 1800°C
3 855 120
ISR 12-LO
MB 25:
900 ... 2500°C
3 855 130
ISR 12-LO
MB 33:
1000 ... 3300°C
3 855 200
IGAR 12-LO MB 10:
300 ... 1000°C
3 855 210
IGAR 12-LO MB 13:
350 ... 1300°C
3 855 220
IGAR 12-LO MB 22:
500 ... 2200°C
Scope of delivery:
Converter, fiber optic, length 2.5 m, optical head I or II, PC software InfraWin, work certificate, user manual.
Note: a connection cable is not included in scope of delivery and has to be ordered separately.
Ordering details:
To process your order as fast as possible, please give us the following data:
Instrument with reference number (e.g. ISR 12-LO; 3 855 100)
Optical head’s design (I or II) and the desired measuring distance (e.g. optical head I, a = 120 mm)
Length of optical fiber (except standard length 2.5 m)
Connecting cable (e.g. 5 m length 3 820 330)