IMPAC pyrometers ISR 12-LO · IGAR 12-LO

8

4.1

Pin assignment for the connector on the bottom side of the pyrometer

Pin

Color

Indication

K

white

+ 24 V power supply (or 24 V AC)

A

brown

0 V power supply

L

green

+ I

outp.

analog output

B

yellow

– I

outp.

analog output

H

gray

external switch for targeting light (bridge to K)

J

pink

see 4.1.1: operating mode contact, external
clearing of maximum value storage (bridge to
K) or output for dirty window monitoring

G

red

DGND (Ground for interface)

F

black

RxD (RS232) or B1 (RS485)

C

violet

TxD (RS232) or A1 (RS485)

D

gray/pink B2 (RS485) (bridge to F)

E

red/blue

A2 (RS485) (bridge to C)

M

orange

Screen only for cable extension,
don’t connect at the switchboard



4.1.1

Connector pin J

The connector pin J can be used for 4 different functions:

1) Operating mode contact: During the warm-up of the pyrometer (after connection to the power supply,

the LED display on the converter indicates “7777”) pin J is connected to the power supply voltage. This
voltage is connected by a relays switch (max. continuous current 0.4 A) to pin K (power supply voltage).
This relays switch opens when the warm-up is finished and the pyrometer is in operating mode. For that
reason this pin can be used as operating mode contact.

2) External clearing of the maximum value storage: When the pyrometer is in operating mode, pin J can

be used for external clearing of maximum value storage (see 9.3). To clear the maximum value storage,
connect pin J for a short time to pin K (power supply voltage).

The function “external clearing” is activated with the following conditions:
• Warm-up is finished (no display “7777“)

• The clear time is set to “extern“ (see 8 and 10.7).

• The lens contamination monitoring is switched off. This can be done via software InfraWin in “color-

bar“ window. The warning level “dirty window” must be set to 0% (see 10.8).

3) Hold function: when the hold function mode is activated the current temperature reading is frozen as

long as J and pin K are connected (see 9.3 clear time for the maximum value storage).

4) Lens contamination monitoring system: The pyrometers are equipped with a contamination monitoring

system. Contamination of lens or sighting window or dust in the sighting path of the pyrometer can inter-
fere the signal in such a high level that a correct temperature measurement will be impossible (this can
also happen if the object is smaller than the spot size of the pyrometer). To avoid wrong measurements in
advance, a warning signal can be set to a certain contamination level. If contamination reaches this level
(or at a certain too low signal level), a built-in relays (max. continuous current 0.4 A) connects pin J to pin
K (power supply voltage). The setting of the switch level (0 … 99%) can be done via software InfraWin in
“color-bar“ window. If the warning level “dirty window” is set to 0% (see 10.8) the lens contamination
monitoring system is switched off and pin J has the function “external clearing” of maximum value storage
(factory setting).

The lens contamination monitoring system is activated with the following conditions:

• Warm-up is finished (no display “7777“)

• The clear time is not set to “extern“ (see 8 and 10.7).

• The pyrometer is operating in „2-color mode“ (see 9.4, 10.7)





K

A

B

C

D

J

G

F

E

H

M

L

Connector

Pin assignment

(side of male inserts)