1 dimensions, Dimensions – LumaSense Technologies ISR 12-LO Benutzerhandbuch
Seite 6

IMPAC pyrometers ISR 12-LO · IGAR 12-LO
5
Maximum value storage:
Built-in single or double storage. Clearing with
adjusted time t
clear
(0.00 s; 0.01 s; 0.05 s; 0.25 s; 1 s; 5 s; 25 s), extern, via
interface or automatically with the next measuring object
Switch contact:
max. 0.15 A (only active with automatic clear mode or t
CL
≥ 0.5 s)
Measuring uncertainty:
(
ε
= 1, t
90
= 1 s, T
U
= 23°C)
ISR 12-LO:
0.4% of reading in °C + 1°C (< 1500°C)
0.6% of reading in °C + 1°C (> 1500°C)
IGAR 12-LO: 0.5% of reading in °C + 1°C (< 1500°C)
0.7% of reading in °C + 1°C (> 1500°C)
Repeatability:
ISR 12-LO:
0.2% of reading in °C + 1°C
IGAR 12-LO: 0.3% of reading in °C + 1°C
Control panel:
4 function keys, switch for serial interface, push button for test current
Protection class:
IP65 (DIN 40 050)
Ambient temperature:
Converter housing: 0 to 60°C (ISR 12-LO); 0 to 50°C (IGAR 12-LO)
Storage temperature:
-20 ... 70°C
Weight converter:
2.2 kg
Sighting:
Laser targeting light (max. power level < 1 mW,
λ = 630-680 nm, CDRH class II).
(With IGAR 12-LO measurements are possible while
laser targeting light is switched on, with ISR 12-LO
measuring is stopped)
CE-label:
According to EU directives about electromagnetic immunity
1.1
Dimensions
Pyrometer housing:
Optical heads:
Type I (small optics):
Type II (focusable optics):
Type II (fixed adjusted):
all dimensions in mm