1 dimensions, Dimensions – LumaSense Technologies IGA 5-LO Benutzerhandbuch
Seite 6

IMPAC pyrometers IS 5 · IS 5-LO · IGA 5 · IGA 5-LO
6
Accuracy:
< 350°C:
0.5% of measured value in °C + 1°C
350 to 1500°C: 0.3% of measured value in °C + 1°C
> 1500°C:
0.5% of measured value in °C + 1°C
Repeatability:
0.2% of measured value in °C + 2°C
Ambient temperature:
0 to 70°C (The laser targeting light switches off if the internal temperature
goes above approx. 50°C. At 70°C a thermal switch is activated
at the 4 to 20 mA output which sets the output to 0 mA)
Storage temperature:
-20 to 70°C
Enclosure rating:
IP65 (DIN 40050)
Weight:
approx. 550 g
Housing:
Stainless steel
Operating position:
Any
CE label:
According to EU directives about electromagnetic immunity
Sighting (optional):
Laser targeting light (max. power level < 1 mW,
λ = 630-680 nm, CDRH class II)
Additional technical data for pyrometers with built-in video module (IS 5-TV; IGA 5-TV)
Video-signal:
composite video-signal approx. 1 V
pp
at 75
Ω, CCIR, 50 Hz
NTSC-standard, EiA, 60 Hz
(optional to order)
Array size:
628 x 583 pixels, b & w
510 x 492 pixels, b & w
Exposure:
automatic, additionally 3-levels controlled by the measuring temperature
Field of view:
approx. 10% x 14% of focused distance
Date/time:
real-time clock with at least 3 days spare run
Video output plug:
separate round plug at the pyrometer, not galvanically separated to the
pyrometer’s power supply
Picture insertions:
target marking; unit number or user text (max. 12 characters),
time and/or date; (individually switchable), measuring temperature,
emissivity
1.1
Dimensions
Instruments:
Optical heads for fibre optic instruments:
Type
Ι
:
Type
ΙΙ
: