Exposure Limit Value (ELV)
The table indicates that the operator using 15 ml Falcon tubes in touch vortex mode (3500 rpm) with reported
usage would reach the ELV in 444 minutes and using 50 ml Falcon tubes in 136 minutes. Based on a usage
pattern of 30 seconds vortex / 5 seconds gap, this means that approximately 761 of the 15 ml or 233 of the 50
ml Falcon tubes could produce exposures that reach the ELV. Immediate action must thus be taken to limit
exposures to below the ELV. With a usage pattern of 15 seconds vortex / 5 seconds gap, 1332 x 15 ml tubes
or 408 x 50 ml tubes must be vortexed until the ELV is reached. With a usage pattern of 45 seconds vortex / 5
seconds gap 532 x 15 ml tubes or 163 x 50 ml tubes must be vortexed.

Number of tubes to reach

Vortex time

(5 sec gap)

EAV ELV

15 ml

EAV ELV

50 ml

15 sec

333

1332

102

408

30 sec

190

761

58

233

45 sec

133

532

40

163

Number of tubes for not exceeding

Vortex

time

(5 sec gap)

EAV ELV

15 ml

EAV ELV

50 ml

15 sec

333

1332

102

408

30 sec

190

761

58

233

45 sec

133

532

40

163