2 Safety notes

- Equipment for potentially explosive areas which contain hybrid

mixtures has to be especially checked for this use. Hybrid mix-
tures are potentially explosive mixtures out of combustible gases,
vapors, or mists with combustible dust. The operator bears the
responsibility of checking if the equipment is suitable for such
uses.

- For transmitters (housing material PC, encapsulation material

PUR) or built-in heads, the compatibility of the instrument material
with the measurement media must be ensured.

- The ambient temperatures must not exceed the limits specified in

the table below.

- The head-mounted temperature transmitter must be set up in

such a way as to achieve at least IP20 enclosure protection as per
EN 60529.

- During setup and operation of the head-mounted temperature

transmitter, care has to be taken that no electrostatic discharge
can occur.