7
Setup interface
The setup interface is available for configuring the transmitter from
a PC. Connection is via the PC interface with USB/TTL converter
and adapter.
Configurable parameters:
• TAG number (10 characters)
• sensor type
• connection circuit (2-/3-/4-wire circuit)
• external or internal cold junction
• range limits
• output signal rising/falling (reversed)
• digital filter
• response on probe break/short-circuit
• recalibration (fine calibration)
• lead resistance for 2-wire circuit
If no supply unit (supply isolator) is available, the 2-wire transmitter
can also be configured using a 9V block battery as a power source.
Please notice that the dTRANS T01 Junior doesn‘t support
custom linearization!
Fine calibration
Fine calibration means the adjustment of the output signal. The sig-
nal can be adjusted within ± 5% of 20mA full scale. Fine calibration
can be performed through the setup program. Values for 4 mA
(zero) and 20 mA (full scale) and offset can be calibrated separately
using the setup program.