The CND/AR measures conductivity using conductance cell and Pt100
for temperature compensation. It displays the conductivity at the
reference temperature, produces a proportional analogue current/
voltage output and has two alarm channels.

Cleaning
The case can be wiped with a damp cloth. De­energise the unit before
cleaning.

Conductance(Ω

-1

),Conductivity(Ω

-1

cm

-1

)andProbefactors

The conductance measurement is the inverse of the resistance
between the two probes. This depends on properties of the probe, like
the distance between them. The probe factor P

f

gives the relationship

between measured conductance M

c

and the actual conductivity

measurement C, so that
C = M

c

x P

f