• Calibration

General
The CND/AR is factory calibrated. Once calibrated it will operate with
any input or output format without recalibration.
Allow the instrument 1 minutes of powered operation (to reach a
stable temperature) before calibration and allow a few seconds for
signals to stabilise before accepting their value.

Equipmentrequirements
• Suitable precision 0 Ω, 200 Ω, 1 kΩ,  kΩ resistors (±0.0 % or

better accuracy)

• An Rtd simulator with ±0.0 % or better accuracy
• An accurate digital multimeter (accurate to 0.0 mV and ±0.1 μA)

Terminalconnectionsforinputcalibration

CalibrationStage

Signaltype

Terminal

Calibrating resistance

2
3

Pt100 RTD simulator

A

4

B

5

Bsense

6

Analogue Current Output

mA output +

14

mA output −

15

Analogue Voltage Output

V output +

14

V output −

16