Terminal

Signal

1

Sense −

Bridge Excitation Voltage

2

Sense +

3

Excitation +

4

Excitation −

6

0 V reference (for calibration only)

Input signal

7

Signal +

8

Signal −

9

Not Used

10

Neutral (−)

Power Supply

11

Live (+)

12

Not Used

13

Output +

Output signal

14

Output −

15,16

Not Used

Case

Earthing is via a stud on lower side of case

Note: The power supply is isolated from the rest of the circuit but input and output circuits share a
common negative return.

− Sense

+ Sense

+ Excitation
− Excitation

+ Signal
− Signal

1 2 3 4

7 8

STG

Input connections for STG with remote sensing excitation.

+ Excitation

− Excitation

+ Signal

− Signal

1 2 3 4

7 8

STG

Input connections for STG with remote sensing disabled.