1

1. Connect up the instrument according to the table above.
. Set the input source to 0.00 mV.
. Adjust the TARE to give 0.00 mV across the test points (multimeter

two). Do not over-adjust.

4. Turn the ZERO adjustment until the output is at ‘zero’.
. Set the input source to ‘full scale’.
6. Adjust the SpAN control until the output is at ‘full scale’.
7. Recheck the ZERO adjustment.

Bridgeexcitationbenchadjustment

To set the excitation voltage for remote sensing mode:
1. Connect up the instrument as above.
. Connect a DVM across the test points marked Ex+ and Ex–.
. Adjust the EX control until the voltage is correct.

•Modifications

Removingthelefthandsideplate

1. Remove the six self tapping screws

marked A-F.

. Remove the two countersunk bolts

marked G & H.

. Remove the left hand side plate and

note the position of the insulator
covering the power supply board.

4. You can now ease the power supply board from the housing.
Warning: take great care when operating AC powered instruments

without the housing as there are high voltages present.

Note: bolt the power supply board and insulator to the side plate

before reassembly.

LEFT HAND SIDE VIEW

G

B

C