IMPAC pyrometers IN 210 · IN 210/4 · IN 210/5

4

1

Technical data

Temperature range:

IN 210:

-32 ... 900°C

(-25.6 ... 1652°F)

IN 210/4: 300 ... 1000°C

(572 ... 1832°F)

IN 210/5: 100 ... 1200°C

(212 ... 2192°F)

Sub range:

User adjustable within the temperature range
(via service interface). Minimum span 51°C

Data handling:

Digital

Spectral range:

IN 200:

8 ... 14 µm

IN 200/4:

3.9 µm

IN 200/5:

5.14 µm

IR detector:

Thermopile

Power supply:

24 V DC ± 25%, ripple < 0.5 V

Power consumption:

max. 0.6 W

Analog output:

Linear current 4 to 20 mA
Note: At ambient temperatures > 75°C the output signal is set to

approx. 3.9 mA

Load:

max. 700

 at 24 V power supply; 400  (18 V); 1000  (30 V)

Service interface:

for parameter programming with portable battery driven setup
device HT 6000

Emissivity ε:

20 to 100% adjustable via service interface

Response time t

90

:

120 ms; up to 10 s adjustable via service interface

Uncertainty:

1% of measured value in °C + 1°C (ε = 1, t

90

= 1 s, T

amb.

= 23°C)

Repeatability:

0.5% of measured value in °C + 1°C (ε = 1, t

90

= 1 s, T

amb.

= 23°C)

Noise Equivalent Tem-
perature Difference
(NETD), (

 = 1, ε = 1):

IN 210:

with t

90

= 120 ms: 0.2°C

(23°C measured temp.)

with

t

90

= 1 s: 0.05°C

(23°C measured temp.)

IN 210/4: with t

90

= 120 ms: 0.6°C

(500°C measured temp.)

with

t

90

= 120 ms: 0.2°C

(1100°C measured temp.)

IN 210/5: with t

90

= 120 ms: 0.15°C

(300°C measured temp.)

with

t

90

= 120 ms: 0.1°C

(1100°C measured temp.)

Ambient temperature:

0 to 70°C

Storage temperature:

-20 to 70 °C

Weight:

approx. 450 g

Housing:

Stainless steel

Operating position:

Any

Protection class:

IP65 (DIN 40 055)

CE-label:

according to EU directives about electromagnetic immunity


1.1 Dimensions

Advertising