IMPAC pyrometers IN 210 · IN 210/4 · IN 210/5

9


Spot sizes for distances which are different from the nominal distance of the selected optics
may be calculated using the following formula:

D

)

D

M

(

a

a

M

2

2

D

)

D

M

(

a

a

M

1

1



Float glass calibration for IN 210/5:


When measuring the temperature of very large and hot surfaces (for example

by the float glass production), additional radiation is received by the pyrometer’s

detector due to unavoidable effects (diffraction, multiple reflection). These

effects increase the temperature output.

To get correct temperature values in this case, the pyrometer must be prepared

ex works. The effect will be compensated by the so-called float glass

calibration and has to be ordered separately if required.





Ø M

1

Ø M

Ø M

2

a

1

a

2

a

Ø D

Optics 1200

Optics 300

Optics 100

IN 210/5

24

43

65

4000

2500

1200

0

15

6

17

32

1000

600

300

0

15

2.5

13

24

300

200

100

0

15

Spot size M [mm]

Measuring distance a [mm]

Spot size M [mm]

Measuring distance a [mm]

Spot size M [mm]

Measuring distance a [mm]