Troubleshooting

- 25 -

Trouble

Possible Causes

Action to be taken

- Tip dripping (instrument leaks)

- Unsuitable tip
- Tip not seated tightly

- Only use high-quality tips
- Press tip on firmly

- Aspirated volume too low

- Pipette shaft cone loose

- Tighten clockwise see

“Handling”, page 8

- Delivered volume too low

- Seal contaminated

- see “Cleaning”, page 19

- Seal damaged

- see “Specifications/Spares”,

page 26

- Non-aqueous solutions pipetted

- see “Adjustment”, page 14

- Piston damaged

- see “Spares”, page 26

- Slow suction or none

- Shaft clogged

- Clean shaft

see “Cleaning”, page 19

- Volume too high

- Pipetting key pressed too far

(into the blow-out position)
before sample take up.

- Operate properly

see “Pipetting”, page 11