WIKA Operating Instructions Pressure Balance High-Pressure Version 2.3

19

3.3 Disassembly

.




After all pressure points have been recorded, ensure that the system is depressurised (see section

3.2.5).




Now the test specimen can be unscrewed from the test specimen connection.


Do not disconnect the test specimen or the piston cylinder system until the
pressure in the pressure balance has been completely released.






The handle bars of the star handle can now be unscrewed from the spindle pump.


3.4 Transport

.

To avoid the piston cylinder system getting damaged during transport (e.g. for
Re-Calibrations), you have to fix the transportation lock at the piston cylinder
system (see figure 1 - 5).

The transport of the instrument should only be carried out on a pallet!




The transportation lock consists of 2 plastic plugs and one metal holder (see figure 1)



Put both plugs in the opposite holes (see figure 2)



Check, if the plugs are in the correct position (see figure 3)



After that push the metal holder carefully over the two plugs (see figure 4). Now the piston cylinder

system is protected during the transport (see figure 5)