Pressure Balance High-Pressure

GB

CPB 5000 HP


WIKA Operating Instructions Pressure Balance High-Pressure Version 2.3

24

5.4 Recalibration

.


The recommended interval between recalibrations is 5 years.

This is the recommendation of the German Calibration Service (DKD)
This interval assumes the system and weights are handled carefully.
If the system is in rough usage, we recommend shortening the interval to about three years.

The pressure balance should be immediately maintained and recalibrated, if:



the operating characteristics deteriorate (duration of free rotation, sink rate, sensitivity)



the weight pieces are damaged or corroded


To avoid the piston cylinder system getting damaged during transport, you have
to fix the transportation lock at the piston cylinder system (see section 3.4)


For recalibration or if you have questions about the optimal recalibration cycle, the DKD lab would be
happy to assist you: