Conditioning, calibration, measurement – Xylem R 503/D Benutzerhandbuch
Seite 15

R 503/P / R 503/D
15
ba15345e02
08/2007
Conditioning, calibration, measurement
General information
When operating the electrode ensure that
!
the filling opening (1) for the bridge electrolyte is open
!
the inner junctions (3) are covered with bridge electrolyte
!
no air bubbles are in the bridge electrolyte
!
the depth of immersion is within the optimum range:
Before measuring
Minimum
depth of
immersion
The ground junction (4) must be
covered
Maximum
depth of
immersion
Approx. 1 cm below the fluid level (2) of
the bridge electrolyte
1
2
3
4
x 500 + R 503/P
1 Before use, condition the double rod combination
electrode according to the operating manual of the
ion sensitive electrode.
2 Remove any air bubbles in the bridge electrolyte by
slightly knocking against the shaft.
3 Calibrate according to the operating manual of the
meter and the analysis specification.