Interferences, Aging – Xylem R 503/D Benutzerhandbuch
Seite 16
Advertising

R 503/P / R 503/D
16
ba15345e02
08/2007
Interferences
The bridge electrolyte ELY/BR/503 contains K
+
and NO
3
-
.
Therefore, it cannot be used for the concentration
determination of potassium, nitrates, perchlorates or
fluoroborates.
Use the following:
!
For NO
3
-
determination: ELY/BR/503/N,
!
For K
+
determination: ELY/BR/503/K
as the bridge electrolyte (see R
Aging
Please note that every electrode undergoes a natural aging
process. The following factors shorten the lifetime
considerably:
!
Incorrect storage
!
Special test samples
!
High temperatures
!
High changes in temperature
The warranty does not cover damage caused by measuring
conditions and mechanical damage.
Advertising
Dieses Handbuch ist für die folgenden Produkte bezogen werden: