1 dimensions, Dimensions – LumaSense Technologies IS 12-TSP Benutzerhandbuch
Seite 6

IMPAC pyrometers IS 12-TSP · IGA 12-TSP
6
address; baud rate; test current output
Additionally (only via interface adjustable):
key lock, wait time, recalibration (with special software)
Emissivity
:
0.100 to 1.000 in 1/1000 steps
Exposure time t
90
:
1 ms (with dynamical adaptation at low signal levels); factory setting 1 s,
adjustable to 0.01 s; 0.05 s; 0.25 s; 1 s; 3 s; 10 s
Maximum value storage:
Built-in single or double storage. Clearing with adjusted time t
clear
(off; 0.01 s;
0.05 s; 0.25 s; 1 s; 5 s; 25 s), extern, via interface or automatically with the
next measuring object
Limit switches:
2 relay outputs (switch-over relay contacts), switch power max. 30 W
(I
max
: 1 A, U
max
: 60 V DC)
Rise and fall times: exposure time of the pyrometer + 2 ms
Control panel:
4 keys, operate with tip of ball-point pen
Protection class:
IP65 (DIN 40 050)
Ambient temperature:
0 to 70°C at housing
Storage temperature:
-20 to 70°C
Relative humidity:
Non condensing conditions
Weight:
2.2 kg
Sighting:
built-in parallax free thru-lens view finder, optionally
with an additional built-in laser targeting light (max.
power level < 1 mW,
= 630-680 nm, CDRH class II)
CE-label:
According to EU directives about electromagnetic immunity
1.1
Dimensions