Jenway Genova Nano Benutzerhandbuch

Seite 15

Advertising
background image

13

4. Wählen Sie die Ziffer, die geändert werden soll, mit den Tasten unter dem Bildschirm aus. Verwenden

Sie die Tasten neben den Pfeil-Symbolen, um die Zahl zu vergrößern oder zu verkleinern. Drücken
Sie auf die Taste neben dem Häkchen-Symbol, um Änderungen zu speichern.

5.2.3

Kalibrierung und überprüfung der Messzelle

Wird bei der Kalibrierung und Überprüfung der Messzelle eine Fehlermeldung angezeigt,
finden Sie weitere Informationen dazu in Kapitel 9.

17

2. Select the digit to be changed using the keys

at the bottom of the screen. Use the keys
adjacent to the arrow icons to increase or
decrease the number. Press the key adjacent
to the tick icon to save any changes.

3. Enter the stated path length (1.0mm for the

supplied calibration standard solution set) for
the DNA standard solution values, in mm, into
the accessory settings by pressing the key
adjacent to the Calibration Standard Path
Length Value icon.

4. Select the digit to be changed using the keys at the bottom of the screen. Use the keys

adjacent to the arrow icons to increase or decrease the number. Press the key adjacent to the
tick icon to save any changes.

5.2.3

Accessory Calibration and Verification

If an error message is displayed during the accessory calibration and verification procedure,
refer to Section 9 for further details.

1. The accessory calibration sequence is initiated

by pressing the key adjacent to the Path
Length Calibration icon.

2. The Air Measurement icon will be displayed.

Use a lint free cloth to clean the upper and
lower read heads, close the instrument lid and
press the key adjacent to the tick icon to
continue. Pressing the key adjacent to the
cross icon will return the instrument to the
micro volume accessory settings menu screen.

17

2. Select the digit to be changed using the keys

at the bottom of the screen. Use the keys
adjacent to the arrow icons to increase or
decrease the number. Press the key adjacent
to the tick icon to save any changes.

3. Enter the stated path length (1.0mm for the

supplied calibration standard solution set) for
the DNA standard solution values, in mm, into
the accessory settings by pressing the key
adjacent to the Calibration Standard Path
Length Value icon.

4. Select the digit to be changed using the keys at the bottom of the screen. Use the keys

adjacent to the arrow icons to increase or decrease the number. Press the key adjacent to the
tick icon to save any changes.

5.2.3

Accessory Calibration and Verification

If an error message is displayed during the accessory calibration and verification procedure,
refer to Section 9 for further details.

1. The accessory calibration sequence is initiated

by pressing the key adjacent to the Path
Length Calibration icon.

2. The Air Measurement icon will be displayed.

Use a lint free cloth to clean the upper and
lower read heads, close the instrument lid and
press the key adjacent to the tick icon to
continue. Pressing the key adjacent to the
cross icon will return the instrument to the
micro volume accessory settings menu screen.

17

2. Select the digit to be changed using the keys

at the bottom of the screen. Use the keys
adjacent to the arrow icons to increase or
decrease the number. Press the key adjacent
to the tick icon to save any changes.

3. Enter the stated path length (1.0mm for the

supplied calibration standard solution set) for
the DNA standard solution values, in mm, into
the accessory settings by pressing the key
adjacent to the Calibration Standard Path
Length Value icon.

4. Select the digit to be changed using the keys at the bottom of the screen. Use the keys

adjacent to the arrow icons to increase or decrease the number. Press the key adjacent to the
tick icon to save any changes.

5.2.3

Accessory Calibration and Verification

If an error message is displayed during the accessory calibration and verification procedure,
refer to Section 9 for further details.

1. The accessory calibration sequence is initiated

by pressing the key adjacent to the Path
Length Calibration icon.

2. The Air Measurement icon will be displayed.

Use a lint free cloth to clean the upper and
lower read heads, close the instrument lid and
press the key adjacent to the tick icon to
continue. Pressing the key adjacent to the
cross icon will return the instrument to the
micro volume accessory settings menu screen.

17

2. Select the digit to be changed using the keys

at the bottom of the screen. Use the keys
adjacent to the arrow icons to increase or
decrease the number. Press the key adjacent
to the tick icon to save any changes.

3. Enter the stated path length (1.0mm for the

supplied calibration standard solution set) for
the DNA standard solution values, in mm, into
the accessory settings by pressing the key
adjacent to the Calibration Standard Path
Length Value icon.

4. Select the digit to be changed using the keys at the bottom of the screen. Use the keys

adjacent to the arrow icons to increase or decrease the number. Press the key adjacent to the
tick icon to save any changes.

5.2.3

Accessory Calibration and Verification

If an error message is displayed during the accessory calibration and verification procedure,
refer to Section 9 for further details.

1. The accessory calibration sequence is initiated

by pressing the key adjacent to the Path
Length Calibration icon.

2. The Air Measurement icon will be displayed.

Use a lint free cloth to clean the upper and
lower read heads, close the instrument lid and
press the key adjacent to the tick icon to
continue. Pressing the key adjacent to the
cross icon will return the instrument to the
micro volume accessory settings menu screen.

2. Wählen Sie die Ziffer, die geändert werden soll, mit

den Tasten unter dem Bildschirm aus. Verwenden
Sie die Tasten neben den Pfeil-Symbolen, um die
Zahl zu vergrößern oder zu verkleinern.

Drücken Sie auf die Taste neben dem Häkchen-
Symbol, um Änderungen zu speichern.

3. Geben Sie die angegebene Schichtdicke (1,0 mm

für die Kalibrierlösung aus dem mitgelieferten
Satz) für die DNA-Standardlösungswerte, in mm,
in die Messzelleneinstellungen ein. Drücken Sie
dazu die Taste neben dem Symbol „Schichtdicke
des Kalibrierstandards“.

1. Starten Sie die Kalibriersequenz für die Messzelle.

Drücken Sie dazu die Taste neben dem Symbol
für die Schichtdickenkalibrierung.

2. Das Symbol „Luftmessung“ wird angezeigt.

Verwenden Sie ein fusselfreies Tuch, um den
oberen bzw. unteren Lesekopf zu reinigen.
Schließen Sie den Gerätedeckel und drücken
Sie auf die Taste neben dem Häkchen-Symbol,
um fortzufahren. Drücken Sie auf die Taste
neben dem Kreuz-Symbol, um wieder zum
Bildschirm mit dem Menü „Einstellungen“ für
die Mikrovolumen-Messzelle zu wechseln.

Das Gerät nimmt fünf Messungen (dunkel bzw.
hell) bei 260 und 330 nm und den Schichtdicken
0,5 und 0,2 mm vor.