guidance and Manufacturer’s declaration-electromagnetic immunity

The Pain Relief Patch BlueTouch is intended for use in the electromagnetic environment specified

below. The customer or the user of the Pain Relief Patch BlueTouch should assure that it is used in

the following environment:
Immunity Test

IEC 60601-1-2

Test Level

Compliance Level

Electromagnetic Environment

- guidance

Electrostatic

discharge (ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be made of

wood, concrete, or ceramic

tile. If floors are covered with

synthetic material, the relative

humidity should be at least 30%.

Electrical fast

transient/burst

IEC 61000-4-4

±2 kV for power

supply lines

±1 kV for input/

output lines

±2 kV for power

supply lines

±1 kV for input/

output lines

Mains power quality should be

that of a typical commercial or

hospital environment.

Surge

IEC 61000-4-5

±1 kV line(s)

to line(s)

±2 kV line(s)

to earth

±1 kV line(s)

to line(s)

±2 kV line(s)

to earth

Mains power quality should be

that of a typical commercial or

hospital environment.

Voltage dips, short

interruptions, and

voltage variations

on power supply

input lines

IEC 61000-4-11

<5% Un

(>95% drop in Un)

for 0,5 cycle

40% Un

(60% drop in Un)

for 5 cycles

70% Un

(30% drop in Un)

for 25 cycles

<5% Un

(>95% drop in Un)

for 5 seconds

<5% Un

(>95% drop in Un)

for 0.5 cycle

40% Un

(60% drop in Un)

for 5 cycles

70% Un

(30% drop in Un)

for 25 cycles

<5% Un

(>95% drop in Un)

for 5 seconds

Mains power quality should be

that of a typical commercial

or hospital environment. If the

user of the Pain Relief Patch

BlueTouch requires continued

operation during power mains

interruptions, it is recommended

that the Pain Relief Patch

BlueTouch be powered from

an uninterruptible power supply

or a battery.

Power frequency

(50/60 Hz)

magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields

should be at levels characteristic

of a typical location in a

typical commercial or

hospital environment.

NOTE Un is the a.c. mains voltage prior to application of the test level